Teledyne T802 Manual do Utilizador Página 189

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 188
Teledyne API T802 Paramagnetic O
2
Analyzer Operation Manual Maintenance Schedule & Procedures
187
07275A (DCN6005)
Table 10-1. T802 Maintenance Schedule
DATE PERFORMED
ITEM ACTION FREQ
CAL
CHECK
REQ’D.
MANUAL
Particulate
Filter
Replace
Weekly or as
needed
No
10.3.1
Verify Test
Functions
Record and
analyze
Weekly or after
any
Maintenance or
Repair
No
11.1.2
Pump
Diaphragm
Replace Annually
Yes
10.3.2
Perform Flow
Check
Check Flow Annually
No
10.3.4
Perform
Leak Check
Verify Leak
Tight
Annually or
after any
Maintenance or
Repair
No
10.3.3
Pneumatic
lines
Examine and
clean
As needed
Yes if
cleaned
Chassis Wipe down As needed
Only if
cover
removed
07275B DCN6418