125
5. In the instrument’s SETUP>MORE>COMM>COM2 menu, make the following
settings:
Baud Rate: 115200
COM2 Mode Settings:
Quiet Mode ON SECURITY MODE OFF
Computer Mode ON MULTIDROP MODE OFF
MODBUS RTU OFF ENABLE MODEM OFF
MODBUS ASCII OFF ERROR CHECKING ON
E,8,1 MODE OFF XON/XOFF HANDSHAKE OFF
E,7,1 MODE OFF HARDWARE HANDSHAKE OFF
RS-485 MODE OFF HARDWARE FIFO ON
COMMAND PROMPT OFF
6. Next, configure your communications software, such as APICOM. Use the
COM port determined in Step 4 and the baud rate set in Step 5. The figures
below show how these parameters would be configured in the Instrument
Properties window in APICOM when configuring a new instrument. See the
APICOM manual (PN 07463) for more details.
Note
USB configuration requires that the baud rates of the instrument
and the PC match; check the PC baud rate and change if needed.
Using the USB port disallows use of the rear panel COM2 port
except for multidrop communication.
07275B DCN6418