Página 1 - PARAMAGNETIC OXYGEN ANALYZER

OPERATION MANUAL MODEL T802 PARAMAGNETIC OXYGEN ANALYZER © TELEDYNE ADVANCED POLLUTION INSTRUMENTATION 9480 CARROLL PARK DRIVE SAN DIEGO,

Página 2

Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual viii CONVENTIONS USED In addition to the safety symbols as presented in the Important

Página 3

Section II Operating Instructions Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 98 5.9.3.6. MANUAL CALIBRATION OF THE ANALOG OUTPUTS

Página 4

Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 99 07275B DCN6418

Página 5 - SAFETY MESSAGES

Section II Operating Instructions Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 100 5.9.3.7. MANUAL ADJUSTMENT OF CURRENT LOOP OPTIO

Página 6 - CONSIGNES DE SÉCURITÉ

Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 101 To adjust the zero and span signal levels of the current outputs, select the ANALO

Página 7 - WARRANTY

Section II Operating Instructions Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 102 current meter (see Figure 3-7 for pin assignment

Página 8

Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 103 5.9.4. TURNING AN ANALOG OUTPUT OVER-RANGE FEATURE ON/OFF In its default configura

Página 9 - ABOUT THIS MANUAL

Section II Operating Instructions Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 104 5.9.5. ADDING A RECORDER OFFSET TO AN ANALOG OUT

Página 10 - REVISION HISTORY

Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 105 5.9.6. SELECTING A TEST CHANNEL FUNCTION FOR OUTPUT A4 The test functions availabl

Página 11 - TABLE OF CONTENTS

Section II Operating Instructions Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 106 SAMPLE RANGE=100.00 % O2=XXX.XX<

Página 12

Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 107 5.9.7. AIN CALIBRATION This is the sub-menu to conduct a calibration of the T802 a

Página 13

ix TABLE OF CONTENTS ABOUT TELEDYNE ADVANCED POLLUTION INSTRUMENTATION (TAPI)...

Página 14 - LIST OF FIGURES

Section II Operating Instructions Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 108 5.9.8. ANALOG INPUTS (XIN1…XIN8) OPTION CONFIGUR

Página 15 - LIST OF TABLES

Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 109 5.10. SETUP MORE  ALRM: USING THE GAS CONCENTRATION ALARMS (OPTION 61) The T802

Página 16 - LIST OF APPENDICES

Section II Operating Instructions Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 110 5.10.1. SETTING THE T802 OPTION 61 CONCENTRATION

Página 17 - PART I

Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 111 This page intentionally left blank. 07275B DCN6418

Página 18

07275B DCN6418

Página 19 - 1.2. FEATURES

113 6. COMMUNICATIONS SETUP AND OPERATION This instrument’s rear panel connections include an Ethernet port, a USB port (option) and two serial co

Página 20 - 1.3. OPTIONS

Section II Operating Instructions Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 114 6.2.1. COM PORT COMMUNICATION MODES Each of the

Página 21

Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 115 Press the following button sequence to select communication modes for one of the C

Página 22

Section II Operating Instructions Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 116 6.2.2. COM PORT BAUD RATE To select the baud rat

Página 23

Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 117 6.2.3. COM PORT TESTING The serial ports can be tested for correct connection and

Página 24 - 2.2.1. SAFETY

Table of Contents Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual x 5.7.1. ID (Machine Identification)...

Página 25 - GETTING STARTED

Section II Operating Instructions Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 118 6.3. REMOTE ACCESS VIA THE ETHERNET For network

Página 26

Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 119 Table 6-3: LAN/Internet Configuration Properties PROPERTY DEFAULT STATE DESCRIPT

Página 27 - 3.2. INSTRUMENT LAYOUT

Section II Operating Instructions Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 120 To view the above properties listed in Table 6-3

Página 28

Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 121 6.3.2. MANUALLY CONFIGURING THE NETWORK IP ADDRESSES There are several circumstan

Página 29

Section II Operating Instructions Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 122 Internet Configuration Touchscreen Functions

Página 30

Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 123 6.3.2.1. CHANGING THE ANALYZER’S HOSTNAME The HOSTNAME is the name by which the a

Página 31

Section II Operating Instructions Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 124 6.4. USB PORT FOR REMOTE ACCESS The analyzer can

Página 32 -  Zero air

Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 125 5. In the instrument’s SETUP>MORE>COMM>COM2 menu, make the following set

Página 33

Section II Operating Instructions Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 126 6.5. COMMUNICATIONS PROTOCOLS This section prese

Página 34 - 3.3. CONNECTIONS AND SETUP

Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 127 Example Read/Write Definition window: Example Connection Setup window: Exa

Página 35

Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual Table of Contents xi 9.4.1. Pressure Calibration...

Página 36 - 2 A3 A4

Section II Operating Instructions Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 128 6.5.2. HESSEN The Hessen protocol is a multidrop

Página 37 - VOLTAGE OUTPUTS

Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 129 6.5.2.2. ACTIVATING HESSEN PROTOCOL Once the COM port has been properly configured

Página 38

Section II Operating Instructions Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 130 6.5.2.3. SELECTING A HESSEN PROTOCOL TYPE Curren

Página 39

Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 131 6.5.2.4. SETTING THE HESSEN PROTOCOL RESPONSE MODE The Teledyne API implementation

Página 40 - Power Supply

Section II Operating Instructions Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 132 6.5.2.5. GAS LIST ENTRY FORMAT AND DEFINITIONS T

Página 41 - “ALARM 1” RELAY

Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 133 6.5.2.6. EDITING OR ADDING HESSEN GAS LIST ENTRIES To add or edit an entry to the

Página 42 - RS-232 CONNECTION

Section II Operating Instructions Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 134 6.5.2.7. DELETING HESSEN GAS LIST ENTRIES To de

Página 43

Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 135 6.5.2.8. SETTING HESSEN PROTOCOL STATUS FLAGS Teledyne API’s implementation of Hes

Página 44

Section II Operating Instructions Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 136 To assign or reset the status flag bit assignmen

Página 45

Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 137 This page intentionally left blank.07275B DCN6418

Página 46

Table of Contents Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual xii 12.1.1. Magnetic Properties of O2 Gas...

Página 47 - RS-232 port

138 7. DATA ACQUISITION SYSTEM (DAS) & APICOM The T802 analyzer contains a flexible and powerful, internal data acquisition system (DAS) that

Página 48 - 3.3.2. PNEUMATIC CONNECTIONS

Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 139 IMPORTANT IMPACT ON READINGS OR DATA DAS operation is suspended whenever its confi

Página 49

Section II Operating Instructions Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 140 Table 7-2: DAS Data Channel Properties PROPERTY

Página 50

Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 141  The CALDAT channel collects data based on events (e.g. a calibration operation)

Página 51

Section II Operating Instructions Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 142 Triggering Events and Data Parameters/Functions

Página 52

Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 143 7.1.3. SETUP DAS VIEW: VIEWING DAS CHANNELS AND INDIVIDUAL RECORDS DAS data and

Página 53

Section II Operating Instructions Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 144 7.1.4. SETUP DAS EDIT: ACCESSING THE DAS EDIT

Página 54 - CALIBRATION

Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 145 7.1.4.1. EDITING DAS DATA CHANNEL NAMES To edit the name of a DAS data channel, fo

Página 55

Section II Operating Instructions Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 146 WARNINGS: Some data may be useful when stored if

Página 56 - 3.4.2. FUNCTIONAL CHECKS

Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 147 diagnostic measurements as well as calibration data such as stability, slope and o

Página 57 - 3.4.3. INITIAL CALIBRATION

Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual Table of Contents xiii Figure 4-1: Front Panel Touchscreen and Display ...

Página 58

Section II Operating Instructions Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 148 07275B DCN6418

Página 59

Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 149 Note When the STORE NUM SAMPLES feature is turned on, the instrument will store t

Página 60

Section II Operating Instructions Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 150 To define the REPORT PERIOD, follow the instruct

Página 61 - SPAN GAS CONCENTRATION

Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 151 7.1.4.5. REPORT PERIODS IN PROGRESS WHEN INSTRUMENT IS POWERED OFF If the instrum

Página 62

Section II Operating Instructions Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 152 To set the NUMBER OF RECORDS, follow the instru

Página 63 - OPERATING INSTRUCTIONS

Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 153 7.1.4.7. RS-232 REPORT FUNCTION The DAS can automatically report data to the commu

Página 64

Section II Operating Instructions Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 154 7.1.4.8. ENABLING / DISABLING THE HOLD OFF FEATU

Página 65 - BASIC OPERATION

Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 155 7.1.4.9. THE COMPACT REPORT FEATURE When enabled, this option avoids unnecessary l

Página 66 - MODE EXPLANATION

Section II Operating Instructions Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 156 7.1.4.11. DISABLING/ENABLING DATA CHANNELS Data

Página 67 - 4.2. SAMPLE MODE

Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 157 7.2. REMOTE DAS CONFIGURATION The DAS can be configured and operated remotely via

Página 68

Table of Contents Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual xiv Table 4-2: Test Functions Defined ...

Página 69 - 4.4. SETUP MODE

Section II Operating Instructions Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 158 Figure 7-3: APICOM User Interface for Configur

Página 70

Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 159 7.2.2. DAS CONFIGURATION VIA TERMINAL EMULATION PROGRAMS Although Teledyne API rec

Página 71 - SETUP MENU

07275A (DCN6005) 160 8. REMOTE OPERATION This section provides information needed when using external digital and serial I/O for remote operation.

Página 72 - CONFIGURATION

Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual Remote Operation 07275A (DCN6005) 161 8.2. INTERACTIVE MODE Interactive mode is used wi

Página 73 - CONCENTRATION

Remote Operation Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 162 07275A (DCN6005) be executed. Press ? <CR> or refer to App

Página 74 - 5.4.3. REPORTING RANGE MODES

Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual Remote Operation 07275A (DCN6005) 163 Text strings: Used to represent data that cannot

Página 75

Remote Operation Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 164 07275A (DCN6005) 8.3. REMOTE ACCESS BY MODEM The T802 can be conn

Página 76

Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual Remote Operation 07275A (DCN6005) 165 To initialize the modem press: SAMPLE RANGE=100.0

Página 77

Remote Operation Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 166 07275A (DCN6005) 8.4. COM PORT PASSWORD SECURITY In order to prov

Página 78

Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual Remote Operation 07275A (DCN6005) 167 This page intentionally left blank. 07275B DCN64

Página 79 - FEATURE

15 PART I – GENERAL INFORMATION AND SETUP 07275B DCN6418

Página 80

07275B DCN6418

Página 81 - MPORTANT

07275A (DCN6005) 169 9. CALIBRATION PROCEDURES This section contains a variety of information regarding the various methods for calibrating a T802

Página 82

Calibration Procedures Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 170 07275A (DCN6005) 9.1. BEFORE CALIBRATION The calibration pr

Página 83 - INTERNAL CLOCK

Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual Calibration Procedures 07275A (DCN6005) 171 Cylinders of calibrated O2 gas traceable to

Página 84

Calibration Procedures Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 172 07275A (DCN6005)  If analog readings are used, the respon

Página 85

Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual Calibration Procedures 07275A (DCN6005) 173 9.2.2. PERFORMING A BASIC MANUAL CALIBRATIO

Página 86 - 5.7.2. INET (ETHERNET)

Calibration Procedures Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 174 07275A (DCN6005) 9.2.3. PERFORMING A BASIC MANUAL CALIBRATI

Página 87

Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual Calibration Procedures 07275A (DCN6005) 175 9.2.3.2. ZERO/SPAN POINT CALIBRATION PROCED

Página 88

Calibration Procedures Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 176 07275A (DCN6005) 9.3. ASSESSING CALIBRATION QUALITY After c

Página 89

Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual Calibration Procedures 07275A (DCN6005) 177 To carry out this adjustment, the current a

Página 90

Section 1 General Information Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 16 This page intentionally left blank.07275B DCN6418

Página 91

Calibration Procedures Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 178 07275A (DCN6005) 9.4.2. FLOW CALIBRATION The flow calibrati

Página 92 - 5.9.1. SIGNAL I/O

Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual Calibration Procedures 07275A (DCN6005) 179 9.5. CALIBRATION OF THE OPTIONAL CO2 SENSO

Página 93 - 5.9.2. ANALOG OUTPUT

Calibration Procedures Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 180 07275A (DCN6005) 9.5.3. ACTIVATE CO2 SENSOR STABILITY FUNC

Página 94 - AIO Configuration Submenu

Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual Calibration Procedures 07275A (DCN6005) 181 (The ACAL submenu in the Primary Setup Menu

Página 95

Calibration Procedures Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 182 07275A (DCN6005) This page intentionally left blank. 0727

Página 96

07275A (DCN6005) 183 PART III – MAINTENANCE AND SERVICE 07275B DCN6418

Página 97

Section III, Technical Information Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 184 07275A (DCN6005) This page intentionally left b

Página 98

07275A (DCN6005) 185 10. MAINTENANCE SCHEDULE & PROCEDURES The T802 Paramagnetic Oxygen Analyzer utilizes a technology that is non-depleting a

Página 99

Section III, Technical Information Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 186 07275A (DCN6005) This page intentionally left b

Página 100 - 07275B DCN6418

Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual Maintenance Schedule & Procedures 187 07275A (DCN6005) Table 10-1. T802 Maintenanc

Página 101 - Analyzer Operation Manual

17 1. INTRODUCTION, FEATURES AND OPTIONS 1.1. T802 OVERVIEW The Model T802 (also referred to as T802) Paramagnetic Oxygen Analyzer is a microproces

Página 102 - - GENERAL SAFETY HAZARD

Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual Maintenance Schedule & Procedures 188 07275A (DCN6005) Table 10-2: T802 Test Funct

Página 103

Maintenance Schedule & Procedures Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 189 10.2. PREDICTIVE DIAGNOSTICS The Test Functio

Página 104

Maintenance Schedule & Procedures Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 190 To change the filter: 1. Turn OFF the analyz

Página 105

Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual Maintenance Schedule & Procedures 191 10.3.3. PERFORMING LEAK CHECKS HAZARD STRO

Página 106

Maintenance Schedule & Procedures Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 192 CAUTION GENERAL SAFETY HAZARD Do not use b

Página 107

Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual Maintenance Schedule & Procedures 193 10.3.5. CLEANING THE OPTICAL BENCH The T802

Página 108 - ENTR accepts the

Maintenance Schedule & Procedures Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 194 This page intentionally left blank. 07275B

Página 109 - 5.9.7. AIN CALIBRATION

Maintenance Schedule & Procedures Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 195 11. TROUBLESHOOTING AND SERVICE This section

Página 110 - Units and whether to display

Troubleshooting and Service Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 196 1. Note any WARNING MESSAGES and take corrective action

Página 111

Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual Troubleshooting and Service 197 Top: the CLR button is available to clear the warnin

Página 113

Introduction, Features and Options Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 18 HAZARD Strong Oxidizer OXYGEN IS A STRONG OX

Página 114

Troubleshooting and Service Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 198 To view or clear the various warning messages press: Su

Página 115 - TESTING

Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual Troubleshooting and Service 199 WARNING MESSAGE FAULT CONDITION POSSIBLE CAUSES RELAY

Página 116

Troubleshooting and Service Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 200 Table 11-2: Test Functions - Indicated Failures TEST F

Página 117

Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual Troubleshooting and Service 201 Figure 11-2: Example of Signal I/O Function (The ACA

Página 118 - 6.2.2. COM PORT BAUD RATE

Troubleshooting and Service Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 202 11.2. USING THE INTERNAL ELECTRONIC STATUS LEDS Several

Página 119 - 6.2.3. COM PORT TESTING

Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual Troubleshooting and Service 203 11.2.2.2. RELAY PCA STATUS LED S D6 (Yellow) O2 Senso

Página 120 - LED FUNCTION

Troubleshooting and Service Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 204 In general, flow problems can be divided into three cat

Página 121

Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual Troubleshooting and Service 205 Figure 11-6: T802 – Internal Pneumatics with CO2 Sen

Página 122

Troubleshooting and Service Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 206 4. If gas flows through the instrument when it is disc

Página 123

Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual Troubleshooting and Service 207 11.3.2.5. ACTUAL FLOW DOES NOT MATCH DISPLAYED FLOW If

Página 124

Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual Introduction, Features and Options 19 Option Option Number Description/Notes Reference

Página 125

Troubleshooting and Service Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 208 4. Disconnect the exhaust line from the optical bench

Página 126

Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual Troubleshooting and Service 209 11.6. SUBSYSTEM CHECKOUT Section 10 discussed a varie

Página 127

Troubleshooting and Service Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 210 Table 11-6: DC Power Supply Acceptable Levels CHECK RE

Página 128 - 6.5.1. MODBUS SETUP

Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual Troubleshooting and Service 211 11.6.6. RELAY BOARD The relay board PCA (04135) can be

Página 129

Troubleshooting and Service Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 212 11.6.9. MOTHERBOARD 11.6.9.1. A/D FUNCTIONS The simple

Página 130 - 6.5.2. HESSEN

Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual Troubleshooting and Service 213 11.6.9.3. STATUS OUTPUTS The procedure below can be us

Página 131

Troubleshooting and Service Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 214 11.6.11. RS-232 COMMUNICATIONS 11.6.11.1. GENERAL RS-23

Página 132

Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual Troubleshooting and Service 215 11.6.12. OPTIONAL CO2 SENSOR There are Two LEDs locate

Página 133

Troubleshooting and Service Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 216 9. After reconnecting the power and pneumatic lines, f

Página 134 - CO2, 0, 111, REPORTED

Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual Troubleshooting and Service 217 2. Turn off power to the instrument, fold down the re

Página 135

Introduction, Features and Options Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 20 Option Option Number Description/Notes Reference

Página 136

Troubleshooting and Service Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 218 QUESTION ANSWER How do I perform a leak check? See Sec

Página 137 - CO2 CONC ALARM 2

Maintenance Schedule & Procedures Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 219 12. PRINCIPLES OF OPERATION The T802 parametri

Página 138

Principles of Operation Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 220 Figure 12-1: Paramagnetic O2 Sensor Design When sample ga

Página 139

Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual Principles of Operation 221 This current is passed to a feedback loop, which generates

Página 140

Principles of Operation Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 222 12.2.1. OPERATION WITHIN THE T802 ANALYZER Operationally, t

Página 141 - 7.1. DAS STRUCTURE

Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual Principles of Operation 223 Figure 12-4: T802 – Internal Pneumatic Flow – Basic Conf

Página 142 - 7.1.2. DEFAULT DAS CHANNELS

Principles of Operation Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 224 Figure 12-5: T802 – Internal Pneumatic Flow with CO2 Sens

Página 143 - IMPORTANT

Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual Principles of Operation 225 As the pressure on the downstream side of the orifice (the

Página 144

Principles of Operation Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 226 compensate the concentration measurement for changes in amb

Página 145 - FUNCTION

Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual Principles of Operation 227 Finally, the CPU issues commands via a series of relays an

Página 146

21 2. SPECIFICATIONS, APPROVALS & COMPLIANCE 2.1. SPECIFICATIONS Table 2-1: T802 Specifications PARAMETER SPECIFICATION O2 Sensor Ranges Min

Página 147

Principles of Operation Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 228 Figure 12-8: CO2 Sensor Option PCA Layout and Electronic

Página 148

Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual Principles of Operation 229 Figure 12-9. CPU Card The CPU includes two types of non-

Página 149

Principles of Operation Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 230 Power Connection for DC Heaters Status LED’s (D2 through D

Página 150

Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual Principles of Operation 231 AC Relay Retainer Plate Retainer Mounting Screws Figur

Página 151

Principles of Operation Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 232 D6 (Yellow) O2 Sensor HeaterD5 (Yellow) –CO2 Sensor Heater

Página 152

Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual Principles of Operation 233 functional block selects a particular signal input (e.g.

Página 153

Principles of Operation Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 234 12.5.6.6. EXTERNAL DIGITAL I/O This External Digital I/O pe

Página 154 - 7.1.4 then press:

Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual Principles of Operation 235 RELAY PCASENSOR SUITESLOGIC DEVICES(e.g. CPU, I2C bus, Tou

Página 155

Principles of Operation Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 236 Figure 12-14: Front Panel and Display Interface Block Dia

Página 156

Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual Principles of Operation 237 12.5.8. SOFTWARE OPERATION The T802 Analyzer is at its hea

Página 157

Specifications, Approvals & Compliance Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 22 2.2. APPROVALS AND CERTIFICATIONS The Te

Página 158

Principles of Operation Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 238 Two conditions must be simultaneously met to switch to the

Página 159

239 13. A PRIMER ON ELECTRO-STATIC DISCHARGE Teledyne API considers the prevention of damage caused by the discharge of static electricity to be extr

Página 160

A Primer on Electro-Static Discharge Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 240 Table 13-1: Static Generation Voltages for Ty

Página 161

Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual A Primer on Electro-Static Discharge 241  Any time a charged surface (including the

Página 162 - REMOTE OPERATION

A Primer on Electro-Static Discharge Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 242  A charge can be induced onto the conductive

Página 163 - 8.2. INTERACTIVE MODE

Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual A Primer on Electro-Static Discharge 243  Simply touching a grounded piece of metal

Página 164

A Primer on Electro-Static Discharge Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 244 13.4.2. BASIC ANTI-ESD PROCEDURES FOR ANALYZER

Página 165

Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual A Primer on Electro-Static Discharge 245 13.4.2.3. TRANSFERRING COMPONENTS FROM RACK T

Página 166 - 8.3. REMOTE ACCESS BY MODEM

A Primer on Electro-Static Discharge Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 246 1. Opening the outer shipping box away from t

Página 167 - 07275A (DCN6005) 165

Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual A Primer on Electro-Static Discharge 247 GLOSSARY Note: Some terms in this glossary ma

Página 168

23 3. GETTING STARTED This section addresses the procedures for unpacking the instrument and inspecting for damage, presents clearance specificatio

Página 169 - 07275A (DCN6005) 167

A Primer on Electro-Static Discharge Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 248 Term Description/Definition DAC Digital-to-Ana

Página 170

Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual A Primer on Electro-Static Discharge 249 Term Description/Definition LED Light Emittin

Página 171 - CALIBRATION PROCEDURES

A Primer on Electro-Static Discharge Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 250 Term Description/Definition RS-485 specificat

Página 172 - 9.1. BEFORE CALIBRATION

Models T802, 802E Appendix A Menu Trees (Reference: 06530C DCN6418) Appendix A A-1 APPENDIX A – Version Specific Software Documentation APPENDIX A

Página 173

Appendix A Models T802, 802E Appendix A Menu Trees (Reference: 06530C DCN6418) A-2 Error! Unknown document property name.Error! Unknown document pr

Página 174

Models T802, 802E Appendix A Menu Trees (Reference: 06530C DCN6418) Appendix A Error! Unknown document property name.Error! Unknown document proper

Página 175

Appendix A Models T802, 802E Appendix A Menu Trees (Reference: 06530C DCN6418) A-4 Error! Unknown document property name.Error! Unknown document pro

Página 176

Models T802, 802E Appendix A Menu Trees (Reference: 06530C DCN6418) Appendix A Error! Unknown document property name.Error! Unknown document proper

Página 177 - 07275A (DCN6005) 175

Appendix A Models T802, 802E Appendix A Menu Trees (Reference: 06530C DCN6418) A-6 Error! Unknown document property name.Error! Unknown document pro

Página 178 - SUBSYSTEMS

Models T802, 802E Appendix A Menu Trees (Reference: 06530C DCN6418) Appendix A Error! Unknown document property name.Error! Unknown document proper

Página 179

Section II Operating Instructions Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 24 Verify that there is no apparent external shippin

Página 180 - 9.4.2. FLOW CALIBRATION

Appendix A Models T802, 802E Appendix A Menu Trees (Reference: 06530C DCN6418) A-8 Error! Unknown document property name.Error! Unknown document pro

Página 181 - SENSOR

Models T802, 802E Appendix A Menu Trees (Reference: 06530C DCN6418) Appendix A Error! Unknown document property name.Error! Unknown document proper

Página 182 - SENSOR STABILITY FUNCTION

Appendix A Models T802, 802E Appendix A Menu Trees (Reference: 06530C DCN6418) A-10 Error! Unknown document property name.Error! Unknown document pr

Página 183 - ZERO/SPAN CALIBRATION:

Models T802, 802E Appendix A Menu Trees (Reference: 06530C DCN6418) Appendix A Error! Unknown document property name.Error! Unknown document proper

Página 184

Appendix A Models T802, 802E Appendix A Menu Trees (Reference: 06530C DCN6418) A-12 Error! Unknown document property name.Error! Unknown document pr

Página 185 - MAINTENANCE AND SERVICE

Models T802, 802E Appendix A Menu Trees (Reference: 06530C DCN6418) Appendix A Error! Unknown document property name.Error! Unknown document proper

Página 186 - 184 07275A (DCN6005)

Appendix A Models T802, 802E Appendix A Menu Trees (Reference: 06530C DCN6418) A-14 Error! Unknown document property name.Error! Unknown document pr

Página 187 - 10.1. MAINTENANCE SCHEDULE

Models T802, 802E Appendix A Menu Trees (Reference: 06530C DCN6418) Appendix A Error! Unknown document property name.Error! Unknown document proper

Página 188 - 186 07275A (DCN6005)

Appendix A Models T802, 802E Appendix A Menu Trees (Reference: 06530C DCN6418) A-16 Error! Unknown document property name.Error! Unknown document pr

Página 189

Models T802, 802E Appendix A Menu Trees (Reference: 06530C DCN6418) Appendix A Error! Unknown document property name.Error! Unknown document proper

Página 190 - ZERO CAL

Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 25 3.1.1. VENTILATION CLEARANCE Whether the analyzer is set up on a bench or installed

Página 191 - 10.3. MAINTENANCE PROCEDURES

Appendix A Models T802, 802E Appendix A Menu Trees (Reference: 06530C DCN6418) A-18 Error! Unknown document property name.Error! Unknown document pr

Página 192

Models T802, 802E Appendix A Menu Trees (Reference: 06530C DCN6418) Appendix A Error! Unknown document property name.Error! Unknown document proper

Página 193 - HAZARD

Appendix A Models T802, 802E Appendix A Menu Trees (Reference: 06530C DCN6418) A-20 Error! Unknown document property name.Error! Unknown document pr

Página 194 - GENERAL SAFETY HAZARD

Models T802, 802E Appendix A Menu Trees (Reference: 06530C DCN6418) Appendix A Error! Unknown document property name.Error! Unknown document proper

Página 195

Appendix A Models T802, 802E Appendix A Menu Trees (Reference: 06530C DCN6418) A-22 Error! Unknown document property name.Error! Unknown document pr

Página 196

Models T802, 802E Appendix A Menu Trees (Reference: 06530C DCN6418) Appendix A Error! Unknown document property name.Error! Unknown document proper

Página 197 - TROUBLESHOOTING AND SERVICE

Appendix A Models T802, 802E Appendix A Menu Trees (Reference: 06530C DCN6418) A-24 Error! Unknown document property name.Error! Unknown document pr

Página 198

Models T802, 802E Appendix A Menu Trees (Reference: 06530C DCN6418) Appendix A Error! Unknown document property name.Error! Unknown document proper

Página 199

Appendix A Models T802, 802E Appendix A Menu Trees (Reference: 06530C DCN6418) A-26 Error! Unknown document property name.Error! Unknown document pr

Página 200

Models T802, 802E Appendix A Menu Trees (Reference: 06530C DCN6418) Appendix A Error! Unknown document property name.Error! Unknown document proper

Página 201

Section II Operating Instructions Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 26 Figure 3-2. Display Screen and Touch Control Th

Página 202

Appendix A Models T802, 802E Appendix A Menu Trees (Reference: 06530C DCN6418) A-28 Error! Unknown document property name.Error! Unknown document pr

Página 203

Models T802, 802E Appendix A Menu Trees (Reference: 06530C DCN6418) Appendix A Error! Unknown document property name.Error! Unknown document proper

Página 204 - CPU Status LED

Appendix A Models T802, 802E Appendix A Menu Trees (Reference: 06530C DCN6418) A-30 Error! Unknown document property name.Error! Unknown document pr

Página 205

Models T802, 802E Appendix A Menu Trees (Reference: 06530C DCN6418) Appendix A Error! Unknown document property name.Error! Unknown document proper

Página 206

Appendix A Models T802, 802E Appendix A Menu Trees (Reference: 06530C DCN6418) A-32 Error! Unknown document property name.Error! Unknown document pr

Página 207

APPENDIX B - Spare Parts Note Use of replacement parts other than those supplied by Teledyne Advanced Pollution Instrumentation (TAPI) may result

Página 208

This page intentionally left blank. B-207275B DCN6418

Página 209 - 11.4. CALIBRATION PROBLEMS

T80XSparePartsList(Ref:072690000ADCN6431,2012April12)PARTNUMBER DESCRIPTION000940700 CD,ORIFICE,.005YELLOW001763500 ASSY,FLOWCTL,110

Página 210

T80XSparePartsList(Ref:072690000ADCN6431,2012April12)OR0000001 ORING,2‐006VT*(KB)OR0000094 ORING,2‐228V,50DUROVI

Página 211 - 11.6. SUBSYSTEM CHECKOUT

M802E without a pump Expendables KitPart Number Description006190600 AKIT, EXP, 802E W/O PUMP, 1UM 009690300 AKIT, TFE FLTR ELEMENT,

Página 212 - 11.6.5. LCD DISPLAY MODULE

Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 27 Table 3-2. Display Screen and Touch Control Description Field Description/Functio

Página 213 - 11.6.7. SENSOR ASSEMBLY

This page intentionally left blank. B-607275B DCN6418

Página 214 - 11.6.9. MOTHERBOARD

Appendix C Warranty/Repair Questionnaire T80X, M80XE (06532C DCN 5798) TELEDYNE API CUSTOMER SERVICE EMAIL: [email protected] PHONE

Página 215 - 11.6.10. CPU

Appendix C Warranty/Repair Questionnaire T80X, M80XE (06532C DCN 5798) TELEDYNE API CUSTOMER SERVICE EMAIL: [email protected] PHONE:

Página 216

APPENDIX D – Wire List and Electronic Schematics 07275B DCN6418D-1

Página 217 - 11.7. REPAIR PROCEDURES

This page intentionally left blank. D-207275B DCN6418

Página 218 - ATTENTION

T80X Interconnect List(Reference: 073800100A DCN6418)Cable PN Signal Assembly PN J/P Pin Assembly PN J/P Pin036490100 CBL ASSY, AC POWERAC Line Power

Página 219 - 11.8. FAQ’S

T80X Interconnect List(Reference: 073800100A DCN6418)Cable PN Signal Assembly PN J/P Pin Assembly PN J/P PinFROM TO066470000 CBL, CO2 & O2 SENSORS

Página 220 - 11.9. TECHNICAL ASSISTANCE

Advanced Pollution InstrumentationA Teledyne Technologies CompanyTELEDYNE'&%$%7 KHLQIRUPDWLRQKHUHRQLVWKHSURSHUW\RI$3,DQGL VVXEPL

Página 221 - PRINCIPLES OF OPERATION

    $%&''&%$7LWOH1XPEHU 5HYLVLRQ6L]H2UFDG%'DWH 0D\ 6KHHWRI)LOH 1?3&%0*5?5-(VHULHV0RWKH

Página 222

    $%&''&%$7LWOH1XPEHU 5HYLVLRQ6L]H2UFDG%'DWH 0D\ 6KHHWRI)LOH 1?3&%0*5?5-(VHULHV0RWKH

Página 223 - 12.2. NDIR MEASUREMENT OF CO

i ABOUT TELEDYNE ADVANCED POLLUTION INSTRUMENTATION (TAPI) Teledyne Advanced Pollution Instrumentation, Inc. (TAPI) is a worldwide market leader in

Página 224 - 12.3. PNEUMATIC OPERATION

Section II Operating Instructions Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 28 Figure 3-3.: Display/Touch Control Screen Mappe

Página 225

    $%&''&%$7LWOH1XPEHU 5HYLVLRQ6L]H2UFDG%'DWH 0D\ 6KHHWRI)LOH 1?3&%0*5?5-(VHULHV0RWKH

Página 226 - 12.4. FLOW RATE CONTROL

                  

Página 227 - 12.4.3. PNEUMATIC SENSORS

    $%&''&%$7LWOH1XPEHU 5HYLVLRQ6L]H2UFDG%'DWH 0D\ 6KHHWRI)LOH 1?3&%0*5?5-(VHULHV0RWKH

Página 228 - 12.5. ELECTRONIC OPERATION

    $%&''&%$7LWOH1XPEHU 5HYLVLRQ6L]H2UFDG%'DWH 0D\ 6KHHWRI)LOH 1?3&%0*5?5-(VHULHV0RWKH

Página 229 - Current

    $%&''&%$7LWOH1XPEHU 5HYLVLRQ6L]H2UFDG%'DWH 0D\ 6KHHWRI)LOH 1?3&%0*5?5-(VHULHV0RWKH

Página 230

    $%&''&%$7LWOH1XPEHU 5HYLVLRQ6L]H2UFDG%'DWH 0D\ 6KHHWRI)LOH 1?3&%0*5?5-(VHULHV0RWKH

Página 231 - 12.5.4. RELAY BOARD

    $%&''&%$7LWOH1XPEHU 5HYLVLRQ6L]H2UFDG%'DWH 0D\ 6KHHWRI)LOH 1?3&%0*5?5-(VHULHV0RWKH

Página 232 - ELECTRICAL SHOCK HAZARD

112233445566D DC CB BA ATitleNumber RevisionSizeBDate: 6/24/2010 Sheet ofFile: N: \PCBMGR\ ..\06696.P1.R3.schdoc Drawn By:GUI Interface06698D

Página 233 - 12.5.4.1. STATUS LEDS

112233445566D DC CB BA ATitleNumber RevisionSizeBDate: 6/24/2010 Sheet ofFile: N:\PCBMGR\..\06696.P2.R3.schdoc Drawn By:06698D2422uHL1Vin5SHDN

Página 234 - 12.5.6. MOTHERBOARD

112233445566D DC CB BA ATitleNumber RevisionSizeBDate: 6/24/2010 Sheet ofFile: N: \PCBMGR\ ..\06696.P3.R3.schdoc Drawn By:06698D4312345To old

Página 235

Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 29 Figure 3-4: Rear Panel Layout 07275B DCN6418

Página 236

112233445566D DC CB BA ATitleNumber RevisionSizeBDate: 6/24/2010 Sheet ofFile: N:\PCBMGR\..\06696.P4.R3.schdoc Drawn By:06698D44TOUCH SCREEN

Página 237 - (when optional CO2

11223344D DC CB BA ATitleNum ber RevisionSizeADate: 5/7/2010 Sheet ofFil e : N:\PCBMGR\..\06882-P1-R0.SchDoc Drawn By:LVDS, Transmitter Boar

Página 238

11223344D DC CB BA ATitleNum ber RevisionSizeADate: 5/6/2011 Sheet ofFil e: N:\PCBMGR\..\06731-1_ETHERNET.SchDocDrawn By:Auxiliary I/O Board

Página 239 - 12.5.9. ADAPTIVE FILTER

11223344D DC CB BA ATitleNum ber RevisionSizeADate: 5/6/2011 Sheet ofFil e: N:\ PCBMGR\..\06731-2_USB.SchDoc Drawn By:Auxiliary I/O Board (

Página 240

11223344D DC CB BA ATitleNum ber RevisionSizeADate: 5/6/2011 Sheet ofFi le: N:\PCBMG R \..\06731-3_ ADC.SchDoc Drawn By:Auxiliary I/O Board

Página 241

Section II Operating Instructions Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 30 Table 3-3. Rear Panel Component Descriptions Com

Página 242 - 10-25% RH

Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 31 Figure 3-5: Internal Layout 07275B DCN6418

Página 243

Section II Operating Instructions Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 32 3.3. CONNECTIONS AND SETUP This section presents

Página 244 - Ground Point

Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 33 3.3.1.2. ANALOG INPUTS (OPTION 64) CONNECTIONS The Analog In connector is used for

Página 245

Section II Operating Instructions Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 34 Channel A4 is special. It can be set by the user

Página 246 - MAINTENANCE

Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 35 Figure 3-8: Current Loop Option Installed CONVERTING CURRENT LOOP ANALOG OUTPU

Página 247

Section II Operating Instructions Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 36 a) Remove the set screw located at the top cente

Página 248 - CAUTION – ESD HAZARD

Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 37 STATUS 1 2 3 4 5 6 7 8 D + SYSTEM OK

Página 249 - GLOSSARY

Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual ii This page intentionally left blank. 07275B DCN6418

Página 250

Section II Operating Instructions Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 38 CONTROL IN A B C D E F

Página 251

Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 39 Figure 3-11: Concentration Alarm Relay Alarm 1 “System OK 2” Alarm 2 “Conc 1” A

Página 252

Section II Operating Instructions Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 40 In this example, O2 Alarm 1 & CO2 Alarm 1 wil

Página 253

Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 41 IMPORTANT IMPACT ON READINGS OR DATA Cables that appear to be compatible because of

Página 254

Section II Operating Instructions Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 42 Figure 3-13. CPU Connector Pin-Outs for RS-232

Página 255

Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 43 ATTENTION COULD DAMAGE INSTRUMENT AND VOID WARRANTY Printed Circuit Assemblies (PCA

Página 256 - SECONDARY SETUP

Section II Operating Instructions Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 44 Figure 3-14: Jumper and Cables for Multidrop Mo

Página 257 - (DAS)

Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 45 Analyzer Analyzer Analyzer Last Analyzer Female DB9Male DB9 RS-232 COM2 RS-232C

Página 258 - COMM / Hessen

Section II Operating Instructions Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 46 RS-485 CONNECTION As delivered from the factory,

Página 259 - PASSDAS RNGE CLK

Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 47 Table 3-8: NISTSRMs Available for Traceability of O2 Calibration Gases NIST-SRM Ty

Página 260

iii SAFETY MESSAGES Important safety messages are provided throughout this manual for the purpose of avoiding personal injury or instrument damage.

Página 261

Section II Operating Instructions Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 48 ATTENTION COULD DAMAGE INSTRUMENT AND VOID WARRAN

Página 262

Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 49 3.3.2.6. CALIBRATION GAS SOURCES The source of calibration gas is also attached to

Página 263

Section II Operating Instructions Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 50 Figure 3-17: T802 Internal Gas Flow (Basic Conf

Página 264

Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 51 Figure 3-18: T802 – Internal Gas Flow with CO2 Sensor Option IMPORTANT Leak Chec

Página 265

Section II Operating Instructions Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 52 3.4. STARTUP, FUNCTIONAL CHECKS, AND INITIAL CALI

Página 266 - “00000000 ”

Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 53 3.4.1.1. WARNING MESSAGES Because internal temperatures and other conditions may be

Página 267

Section II Operating Instructions Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 54 Table 3-10: Warning Messages MESSAGE MEANING AN

Página 268

Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 55 make sure that the DHCP has successfully downloaded the appropriate network setting

Página 269

Section II Operating Instructions Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 56 of the instrument to finish readying the analyzer

Página 270

Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 57 Verify/change these settings by pressing: 3.4.3.1. REPORTING RANGE SETTINGS 07275B

Página 271

Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual iv CONSIGNES DE SÉCURITÉ Des consignes de sécurité importantes sont fournies tout au lo

Página 272

Section II Operating Instructions Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 58 3.4.3.2. DILUTION RATIO (OPTION) SET UP If the di

Página 273 - Description

Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 59 3.4.3.3. SET O2 SPAN GAS CONCENTRATION Set the expected O2 span gas concentration.

Página 274

Section II Operating Instructions Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 60 3.4.3.4. ZERO/SPAN CALIBRATION To perform the ze

Página 275

Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 61 SECTION II – OPERATING INSTRUCTIONS 07275B DCN6418

Página 276

Section II Operating Instructions Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 62 This page intentionally left blank.07275B DCN6418

Página 277 - Name Description Units

63 4. BASIC OPERATION The T802 analyzer is a computer-controlled analyzer with a dynamic menu interface that allows all major operations to be cont

Página 278

Basic Operation Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 64 Figure 4-1: Front Panel Touchscreen and Display The mode field of

Página 279

Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual Basic Operation 65 4.2. SAMPLE MODE This is the analyzer’s standard operating mode. I

Página 280

Basic Operation Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 66 Table 4-2: Test Functions Defined PARAMETER DISPLAY TITLE UNITS

Página 281

Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual Basic Operation 67 4.4. SETUP MODE The SETUP mode contains a variety of choices that

Página 282

v WARRANTY WARRANTY POLICY (02024F) Teledyne Advanced Pollution Instrumentation (TAPI), a business unit of Teledyne Instruments, Inc., provides tha

Página 283

Basic Operation Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 68 4.4.2. SECONDARY SETUP MENU (SETUP>MORE) Table 6-5: Secondary S

Página 284

69 5. SETUP MENU The SETUP menu is used to set instrument parameters for performing configuration, calibration, reporting and diagnostics operation

Página 285 - APPENDIX B - Spare Parts

Section II Operating Instructions Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 70 5.4. SETUP  RNGE: ANALOG OUTPUT REPORTING RANGE

Página 286

Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 71 5.4.2. ANALOG OUTPUT RANGES FOR O2 CONCENTRATION The analyzer has several active an

Página 287 - T80XSparePartsList

Section II Operating Instructions Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 72 5.4.3. REPORTING RANGE MODES The T802 provides th

Página 288

Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 73 5.4.3.1. RNGE  MODE  SNGL: CONFIGURING THE T802 ANALYZER FOR SINGLE RANGE MODE S

Página 289

Section II Operating Instructions Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 74 5.4.3.2. RNGE  MODE  DUAL: CONFIGURING THE T802

Página 290

Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 75 Note If the optional CO2 sensor is installed, the concentration field of the T802’s

Página 291 - T80X, M80XE

Section II Operating Instructions Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 76 5.4.3.3. RNGE  MODE  AUTO: CONFIGURING THE T802

Página 292

Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 77 (The ACAL submenu in the Primary Setup Menu is a special configuration; consult fa

Página 293

Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual vi This page intentionally left blank. 07275B DCN6418

Página 294

Section II Operating Instructions Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 78 3. Ensure that the upper span limit entered for

Página 295 - T80X Interconnect List

Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 79 password (default 818) is required to enter the VARS or DIAG menus in the SETUP>

Página 296

Section II Operating Instructions Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 80 Example: If all passwords are enabled, the follow

Página 297 - TELEDYNE

Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 81 5.6. SETUP  CLK: SETTING THE T802 ANALYZER’S INTERNAL CLOCK The analyzer has an in

Página 298

Section II Operating Instructions Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 82 5.6.2. ADJUSTING THE INTERNAL CLOCK’S SPEED In or

Página 299

Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 83 5.7. SETUP  MORE COMM: COMMUNICATION PORTS This section introduces the communica

Página 300

Section II Operating Instructions Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 84 To edit the instrument’s ID code, press: (The AC

Página 301

Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 85 5.7.3. COM1[COM2] (MODE, BAUDE RATE AND TEST PORT) Use the SETUP>COMM>COM1[CO

Página 302

Section II Operating Instructions Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 86 Table 5-2: Variable Names (VARS) NO. VARIABLE D

Página 303

Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 87 To access and navigate the VARS menu, use the following key sequence. SETUP X.X

Página 304

vii ABOUT THIS MANUAL This operation manual, PN 07275, is comprised of multiple documents in PDF format, as listed below. Part No. Rev Name/Descr

Página 305

Section II Operating Instructions Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 88 5.9. SETUP  MORE  DIAG: DIAGNOSTICS FUNCTIONS

Página 306

Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 89 To access the DIAG functions press the following menu sequence: SAMPLE RANGE=100.00

Página 307 - GUI Interface

Section II Operating Instructions Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 90 5.9.1. SIGNAL I/O The signal I/O diagnostic mode

Página 308

Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 91 5.9.2. ANALOG OUTPUT The T802 analyzer comes equipped with four analog outputs. 

Página 309 - MBUS B B

Section II Operating Instructions Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 92 To access the ANALOG I/O CONFIGURATION sub menu,

Página 310

Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 93 5.9.3.1. ANALOG OUTPUT VOLTAGE / CURRENT RANGE SELECTION In its standard configurat

Página 311 - To LCD Display

Section II Operating Instructions Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 94 5.9.3.2. CALIBRATION OF THE ANALOG OUTPUTS Analo

Página 312 - DCN:6092

Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 95 5.9.3.3. ENABLING OR DISABLING THE AUTOCAL FOR AN INDIVIDUAL ANALOG OUTPUT To enabl

Página 313 - Auxiliary I/O Board (USB)

Section II Operating Instructions Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 96 5.9.3.4. AUTOMATIC GROUP CALIBRATION OF THE ANALO

Página 314 - Auxiliary I/O Board (ADC)

Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual 97 5.9.3.5. AUTOMATIC INDIVIDUAL CALIBRATION OF THE ANALOG OUTPUTS To use the AUTO CAL