Teledyne M201E Manual do Utilizador Página 33

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 51
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 32
Teledyne API Model M201E NH
3
Analyzer Operator Manual Configurable Analog Output Overview
05206H DCN5910 31
To calibrate the output channels as a group press the following, (Note: this is the
recommended calibration approach):
DIAG AIO AUTO CALIBRATING DATA_OUT_1
AUTO CALIBRATING DATA_OUT_2
AUTO CALIBRATING DATA
_
OUT
_
4
Exit at Any
Time to Return
to the main
DIAG Menu
DIAG
A
NALOG I / O CONFIGURATION
PREV NE XT ENTR EXIT
DIAG AIO
A
OUTS CALIBRATED: NO
< SET SET> CAL EXIT
If AutoCal has been manually
t urned off for any cha nnel the
message f or that c hannel will be
similar to:
NOT AUTO CAL DATA_OUT_1
Exit to Return
to the I/O
Configuration
Menu
FROM ANALOG I/O CONFIGURATION MENU
DIAG AIO AOUTS CALIBRATED: YES
< SET SET> CAL EXIT
Figure 3-4. Analog Output Group Calibration
Vista de página 32
1 2 ... 28 29 30 31 32 33 34 35 36 37 38 ... 50 51

Comentários a estes Manuais

Sem comentários