Getting Started Model 360E Instruction Manual
3.3.1. Initial O
2
Sensor Calibration Procedure
If your instrument has an O
2
sensor option installed that should be calibrated as well.
3.3.1.1. O
2
Calibration Setup
The pneumatic connections for calibrating are as follows:
Figure 3-13: O
2
Sensor Calibration Set Up
O
2
SENSOR ZERO GAS: Teledyne Instruments’ recommends using pure N
2
when calibration the zero point of
your O
2
sensor option.
O
2
SENSOR SPAN GAS: Teledyne Instruments’ recommends using 21% O
2
in N
2
when calibration the span
point of your O
2
sensor option.
3.3.1.2. O
2
Calibration Method
STEP 1 – SET O
2
SPAN GAS CONCENTRATION :
Set the expected O
2
span gas concentration.
This should be equal to the percent concentration of the O
2
span gas of the selected reporting range (default
factory setting = 20.8%; the approximate O
2
content of ambient air).
SAMPLE GAS TO CAL:CO2
CO2 O2 ENTR EXIT
SAMPLE RANGE = 500.000 PPM CO2 =XXX.X
< TST TST > CAL SETUP
M-P CAL A1:NXCNC1 =100PPM NOX=X.XXX
<TST TST> ZERO SPAN CONC EXIT
SAMPLE GAS TO CAL:O2
NOX O2 ENTR EXIT
M-P CAL O2 SPAN CONC:20.8%
020.80 ENTREXIT
The O
X
span concentration value automatically defaults to
20.8 %.
If this is not the the concentration of the span gas being
used, toggle these buttons to set the correct concentration
of the O
2
calibration gases.
EXIT ignores the new
setting and returns to
the previous display.
ENTR accepts the new
setting and returns to
the previous menu.
30 05232 Rev B3