Teledyne 360E Manual do Utilizador Página 128

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 282
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 127
Operating Instructions Model 360E Instruction Manual
6.13.4.5. AIN Calibration
This is the sub-menu to conduct the analog input calibration. This calibration should only be necessary after
major repair such as a replacement of CPU, motherboard or power supplies. Activate the ANALOG I/O
CONFIGURATION MENU (See Section 6.13.1), then press:
Exit at any time to
return to the main
DIAG menu
DIAG
A
NALOG I / O CONFIGURATION
PREV NEXT ENTR EXIT
Continue pressing SET? until …
DIAG AIO
A
IN CALIBRATED: NO
< SET SET> CAL EXIT
DIAG AIO AIN CALIBRATED: YES
< SET SET> CAL EXIT
Instrument
calibrates
automatically
Exit to return to the
ANALOG I/O
CONFIGURATION
MENU
STARTING FROM ANALOG I / O CONFIGURATION MENU
DIAG AIO CALIBRATING A/D ZERO
CALIBRATING A/D SPAN
116 05232 Rev B3
Vista de página 127
1 2 ... 123 124 125 126 127 128 129 130 131 132 133 ... 281 282

Comentários a estes Manuais

Sem comentários