Teledyne T101 Manual do Utilizador Página 158

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 157
Remote Operation Teledyne API – T101 Operation Manual
156
5.1.4.3. Activating Hessen Protocol
The first step in configuring the T101 to operate over a Hessen protocol network is to
activate the Hessen mode for COMM ports and configure the communication parameters
for the port(s) appropriately. Press:
SAMPLE ENTER SETUP PASS : 818
8 1 8 ENTR EX IT
SETUP X.X PRIMARY SETUP MENU
CFG DAS RNGE PASS CLK MORE EXIT
SAMPLE RANGE = 500.0 PPB H2S =XXX.X
< TST TST > CAL SETUP
SETUP X.X SECON DARY SETUP MENU
COMM VARS DIAG ALRM EXIT
SETUP X.X COMMUNICATIONS MENU
ID INET COM1 COM2
EXIT
SETUP X. X COM1 MODE:
0
SET> EDIT EXIT
Co ntinue pr essin g ne xt un til
Select which COMM
port to configure
SETUP X.X COM1 QUIET MODE: OFF
NEXT OFF ENTR EXIT
SETUP X. X COM1 HESSEN PROTOCOL : ON
PRE V NEXT ON ENTR EXIT
SETUP X.X COM1 HESSEN PROTOCOL : OFF
PREV NE XT OFF ENTR EXIT
ENTR accepts the new
settings
EXIT ignores the new
settin
g
s
The sum of the mode
IDs of the selected
modes is displayed
here
SE TUP X. X COM1 E,7,1 MODE: ON
PREV NEXT ON ENTR EXIT
SETUP X.X COM1 E,7,1 MODE: OFF
PREV NEXT OFF ENTR EXIT
Toggle OFF/ON to
change
activate/deactivate
selected mode.
Repeat the
entire process to
set up the
COM2 port
07266B DCN6485
Vista de página 157
1 2 ... 153 154 155 156 157 158 159 160 161 162 163 ... 370 371

Comentários a estes Manuais

Sem comentários