Model GFC7000E Instruction Manual Calibration Procedures
04584 Rev A1 141
To program the Sequence:
Toggle keys
to set
number of
days
between
procedures
(1-367)
Toggle keys
to set
delay time for
each iteration
of the
sequence:
HH:MM
(0 – 24:00)
SETUP X.X
DELTA DAYS: 1
<SET SET> EDIT EXIT
SETUP X.X
DELTA DAYS: 1
0 0 2 ENTR EXIT
SETUP X.X
DELTA DAYS:2
<SET SET> EDIT EXIT
SETUP X.X DELTA TIME00:00
<SET SET>
EDIT
EXIT
SETUP X.X DELTA TIME: 00:00
0 0 :3 0 ENTR EXIT
SETUP X.X DELTA TIEM:00:30
<SET
SET>
EDIT EXIT
SETUP X.X
STARTING TIME:14:15
<SET SET> EDIT EXIT
EXIT returns
to the SETUP
Menu
SETUP X.X DURATION:15.0 MINUTES
<SET SET> EDIT EXIT
SETUP X.X DURATION 15.0MINUTES
3 0 .0 ENTR EXIT
SETUP X.X DURATION:30.0 MINUTES
<SET SET> EDIT EXIT
Toggle keys
to set
duration for
each
iteration of
the
sequence:
Set in
Decimal
minutes
from
0.1 – 60.0
SETUP X.X CALIBRATE: OFF
<SET SET>
EDIT
EXIT
SETUP X.X CALIBRATE: OFF
ON
ENTR
EXIT
SETUP X.X CALIBRATE: ON
<SET SET> EDIT
EXIT
Toggle key
between
Off and
ON
SETUP X.X SEQ 2) ZERO
SPAN, 2:00:30
PREV NEXT MODE SET
EXIT
Sequence Delta Time
MODE Delta Days
SAMPLE
ENTER SETUP PASS : 818
8 1 8 ENTR EXIT
SETUP X.X PRIMARY SETUP MENU
CFG
ACAL
DAS RNGE PASS CLK MORE EXIT
SETUP X.X SEQ 1) DISABLED
NEXT MODE EXIT
SETUP X.X SEQ 2) DISABLED
PREV NEXT
MODE
EXIT
SETUP X.X MODE: DISABLED
NEXT ENTR EXIT
SETUP X.X
MODE: ZERO
PREV NEXT ENTR EXIT
SETUP X.X MODE: ZERO
SPAN
PREV NEXT ENTR EXIT
SETUP X.X SEQ 2) ZERO
SPAN, 1:00:00
PREV NEXT MODE
SET
EXIT
SETUP X.X TIMER ENABLE: ON
SET>
EDIT EXIT
SETUP X.X
STARTING DATE: 01
JAN
02
<SET SET> EDIT EXIT
SETUP X.X
STARTING DATE: 01
JAN
02
0 4 SEP 0 3 ENTR EXIT
Toggle keys
to set
day, month &
year:
Format :
DD-MON-Y
Y
SETUP X.X STARTING DATE: 04
SEP
03
<SET SET> EDIT EXIT
Default
value is
ON
SETUP X.X STARTING DATE: 04
SEP
03
<SET
SET>
EDIT EXIT
SETUP X.X STARTING TIME:00:00
<SET SET>
EDIT
EXIT
SETUP X.X
STARTING TIME:00:00
1 4 : 1 5 ENTR EXIT/
Toggle keys to
set time:
Format : HH:MM
This is a 24 hr
clock .
PM hours are
13 – 24.
Example
2:15 PM = 14:15
SAMPLE* RANGE = 500.000 PPM CO2 =X.XXX
< TST TST > CAL CALZ CALS
SETUP